TECHNICAL SPECIFICATIONS |
|
Frequency |
Tuning range |
9 kHz - 26.5 GHz |
|
Tuning
resolution |
minimum 1 Hz |
|
Frequency
span width |
100 Hz/div
to 300MHz/div:1,2,5 step selections (auto
selected) plus ZERO Span, and FULL Span
(9kHz to 26.5GHz).
Start, Stop, span-manual selections |
|
Span
accuracy |
±3%
of the indicated Span Width, |
|
Readout
accuracy |
Span accuracy + Frequency
Standard Accuracy + 50% of RBW. |
|
Frequency
Counter |
1 kHz, 100 Hz, 10 Hz, 1 Hz |
|
Resolution
Accuracy |
(Reference frequency error
+ counter resolution ± 1count) |
|
Sensitivity |
≤ -70dBm (50kHz ~30GHz) |
|
Stability
Residual FM |
≤
100Hzp-p in 200ms @1kHz, RBW, 1kHz VBW
(P-P in 20ms) |
|
Noise
Sidebands |
-90dBc/Hz @ 10kHz offset |
|
Amplitude |
Measurement Range
Average Displayed
Noise Level |
+30dBm to displayed average
noise level |
|
Pre-amp ON |
Pre-amp OFF |
50kHz to 50 MHz :
-95~-115dBm
50kHz to 1.8 GHz :
-130dBm
1.8GHz to 3.0GHz :
-129dBm |
50kHz to 100 MHz :
-105dBm
100kHz to 2.8 GHz :
-110dBm
2.8GHz to 3.0GHz :
-105dBm
3.0GHz to 26.5GHz :
-110dBm |
| |
1 dB
Compression Point |
(1kHz RBW, 10Hz Video
Filter)
-10dBm, 100kHz to 3.0GHz(0dB attenuation) |
| |
Display
Scale |
100dBm in 10dB/div log
scale.
50dBm in 5dB/div log scale.
20dBm in 2dB/div log scale.
10dBm in 10dB/div log scale.
10 divisions with liner scale. |
| |
Amplitude
Units
Log Scale Mode
Linear Scale Mode
Quasi Peak Enabled |
dBmV or dBm units.
V(uV, mV, etc.), or dBV(dBmV)
dBuV, dBmV, O dBm, |
| |
Display
Linearity |
5 or 10dB/div <± 1.0 dB
over 10 divisions.
1 or 2dB/div <±
0.5 dB over 10 divisions.
Linear, 10% of Refernce Level over 10
divisions |
| |
Frequency
Response |
≤ -3.0 ~ + 1 dB, 9kHz to
5MHz
≤
±1.0dB, 5MHz to 2.9GHz
≤
±1.5dB, 3.0GHz to 6.4GHz
≤
±2.2dB, 6.4GHz to 26.5GHz |
| |
Attenuator
Range
Resolution
Accuracy |
0 to 55 dB Selected
manually or automatically coupled to
reference level.
5 dB steps.
0.5dB/step up to 1.5dB maxinum to 100MHz |
| |
Reference
Level
Range
Resolution
Accuracy |
±1.5dB
(50kHz to 26.5GHz)
-110dBm to + 30dBm with 1 KHz RBW USING 1
dB/div scale
0.1dB step minimum |
| |
Residual
Spurious |
≤
-85dBm (Input terminated, 0 dB attenuation) |
| |
Harmonic
Distortion |
≤
-65dBc, -30dBm Input, 0dB attenuation. |
| |
Intermodulation Distortion |
-70dBc, 100MHz to 26.5GHz,
-30dBm input (3rd order) 0dB Attenuation
-65dBc, 1MHz to 100MHz,
-30dBm input |
| |
Other Input
Related Spurious |
-60dBc, 10MHz to 26.5GHz,
-30dBm input |
| |
Resolution
Bandwidth
Selections |
300Hz, 1kHz, 3kHz, 10kHz,
30kHz, 100kHz,
300kHz, 1MHz, 3MHz
9k, 120kHz : Quasi-Peak Detection : (option
QP-01)
10Hz, 30Hz, 100Hz : D-RBW : (option RB-01) |
| |
Accuracy |
±20% |
| |
Selectivity |
60dB/3dB ratio <15:1
60dB/6dB ratio <12:1;9kHz & 120kHz (option
QP-01) |
| |
Switching
Error |
±1.0dB (3kHz Reference RBW) |
| |
Video
Bandwidth
Selection |
10Hz to 1MHz in 1-3-10
steps plus None. |
|
Sweep |
Time |
20ms to 1000sec
5ms to 15sec (ZERO SPAN)
25us to 2ms (Fast ZERO SPAN option) |
| |
Sweep Time
Accuracy |
<1% (DDS MODE, Span ≤ 10MHz)
<10%(Analog MODE, Span
≤ 10MHz) |
| |
Trigger |
|
| |
Source |
External (rear), Line,
Video, Free run, TV trig |
| |
Modes |
Continuous, Single. |
| |
Coupling |
DC |
| |
External
Level (Rear) |
TTL Level |
| |
Delay |
±1 sweep time (zero span) |
|
Memory |
Trace
Storage |
Maximum 1,000 traces. (*.TRC) |
| |
Setup
Storage |
Maximum 2,000 traces. (*.STS) |
| |
Image
Storage |
30 ~ 200 storage (*.BMP, *.JPG) |
|
Display |
Type |
6.4" Color
TFT LCD |
| |
Digital
resolution |
640H X 480V active display
area |
| |
Marker
Modes |
Peak search, Peak Track,
Delta Marker, 1/Delta
Marker to Reference ( 9 Markers maximum ) |
|
RF Input |
Connector |
N Type Female, 50 Ohm
nominal. |
| |
VSWR |
150kHz to 3.0GHz, VSWR<1.5:1.(with
10dB Input attenuation) |
| |
Maximum
Input |
±50VDC +30dBm (with 10dB
attenuation) |
| |
LO
Emissions |
-70dBm (with 0dB
attenuation) |
|
Output |
IF Output |
10.7MHz |
| |
Video
Output |
0~5VDC (TTL Level) |
| |
Swp Gate |
0~5VDC (TTL Level) |
| |
VGA Out |
External VGA Output (Color
Output) |
| |
Probe Power |
3Pin (+15V, -12V, GND)
connector |
|
AM
Demodulation |
Demodulation Range |
5% to 80% @ 1kHz, 50%
modulation, -20dBm Input |
| |
Input Level
Range |
-2.0dBm to -75dBm @ 1kHz,
50% modulation, |
| |
Frequency
Response |
20Hz to 30kHz @ -20dBm
Input |
| |
Distortion |
≤5% @ 90% modulation @
1kHz, -20dBm Input
≤2% @ 50% modulation @ 1kHz, -20dBm
Input |
| |
Division |
25% / div |
|
FM
Demodulation |
Demodulation Range |
≤
80kHz |
| |
Input Level
Range |
-2.0dBm to -75dBm @ 50kHz
deviation |
| |
Frequency
Response |
20Hz to 100kHz @ -20dBm
Input |
| |
Distortion |
≤5%
@ 20kHz deviation @ 1kHz, -20dBm Input
≤2% @ 50kHz deviation @ 1kHz, -20dBm Input |
| |
Division |
25% / div |
|
Quasi Peak
Detector (option QP-01) |
Band B
Frequency Range |
0.15MHz to 30MHz |
| |
Charge Time |
1ms |
| |
Discharge
Time |
160ms |
| |
Display
Time |
160ms |
| |
Band C
Frequency Range |
30MHz to 100MHz |
| |
Charge Time |
1ms |
| |
Discharge
Time |
550ms |
| |
Display
Time |
100ms |
Frequency
STANDARD
(10Mhz,Ref.) |
|
Standard |
High Stability (option HS-01) |
| |
Temperature
Stability |
±
2ppm |
±
0.2ppm |
| |
Aging |
±
1ppm/year |
±
0.1ppm/year |
| |
Connector |
BNC female connector. |
| |
Input level |
5dBm to + 15 dBm |
| |
Output
level |
+5dBm nominal |
|
IEEE-488
(GPIB)
interface |
Conforms to
|
IEED-Standards 488.1 -
1987. 488.2 - 1992. |
| |
Implemented
subsets |
SH1, AH1, T5, L3, SR1, RL1,
PPO, DC1, DT1, CO, PPO
SR0, DC0, DT0, C1, C2, C3 |
|
RS-232C
interface |
Type |
Null Modem |
| |
Baud Rate |
600bps, 1200bps, 2400bps,
4800bps, 9600bps, 19.2kbps
38.4kbps, 57.6kbps, 115.2kbps |
| |
Parity
Check |
Odd, Even or None, Mark,
Space |
| |
Data Length |
7its, 8bits. |
| |
Stop Bits |
1bit, 2bits |
| |
Protocol |
None, Xon-Xoff, RTS-CTS,
DTR-DSR |
|
Printer |
Drivers |
PCL3 or higher |
| |
Connection |
Standard 25Pin female D-Sub
parallel printer |
|
Power probe |
Power |
HP 85024A compatible
3 pins (+15V, -12V, GND) |
|
General
characteristics |
Dimensions |
350mm : Width (including
handle)
185mm :
Height
381mm : Length |
| |
Weight |
9.4kg |
| |
Warm-up
Time |
15 minutes for the accurate
measurement |
| |
Power
requirements |
|
| |
Source
Voltage & Frequency |
90-250VAC 50/60 Hz |
| |
Power
Consumption |
90Watts maximum (without
options) |
| |
Fuse (F1
and F2) |
3.15A, 250V, Type T |
| |
Environmental Specifications |
|
| |
Place |
Indoors |
| |
Operation
Temperature |
0 to 40°C |
| |
Storage
Temperature |
-20 to 70°C |
| |
Temperature
/ Humidity |
MIL -T-28800E : Type 2,
Class 5 (Operating : 85, Storage : 90%) |
| |
Vibration /
Shock |
MIL -T-28800 : Type 2,
Class 5 |
| |
Altitude |
Up to 3,000 meters (operation)
Up to 40,000feet (none-operation) |
| |
Safety
specifications |
EN61010-1 |
| |
Main supply
voltage fluctuations |
Norminal voltage 10% |
| |
Transient
overvoltage |
Installation Category II |
| |
Pollution
degree |
2 |
| |
RF
emissions and Immunity |
|
| |
RF
emissions |
EN55011 : 1991, Class A |
| |
RF Immunity |
EN50082-1 : 1997 |